Certificate of Calibration

Mass Artifact Submitted By:
The Scale People Inc-MD
9693-C Gerwig Lane
Columbia, MD 21046
Mass Artifact Information:
Date of Calibration: Next Calibration: Purchase Order:
29-Jun-2023 30-Jun-2024 n/a
Description of Mass Artifact:
Manufacturer: Asset ID: Accuracy Class: Description:
Unknown n/a OIML E2 1mg – 500g Weight Kit
Environmental Condition Range:
Temperature: Barometric Pressure: Relative Humidity: Air Density:
20 °C to 21.9 °C 1004 mbar to 1005 mbar 42% to 44% See below
Visual Inspection of Mass Artifact when Received:
Good
Notes Regarding Calibration:
Cleaned and Calibrated
Procedure Used in Calibration:
SOP 32 – Double Substitution

This calibration certificate shall not be reproduced except in full and without written permission of The Lab People, Inc.

Certificate of Calibration
Nominal ID True Mass (g) True Mass Corr. (mg) Conv Mass (g) Conv Mass Corr (mg) (k=2) Unc (mg) MPE (mg) MPE
P/F
Assumed Density
g/cm3
Balance Used Reference Standard Air Density
1 mg 1M5GK2 0.0010035 0.0035 0.0010032 0.0032 0.0013 0.006 P 2.7 XPR6U 9VKV 0.001185
2 mg 1M5GK2 0.0020006 0.0006 0.0020000 0.0000 0.0014 0.006 P 2.7 XPR6U 9VKV 0.001185
2 mg* 1M5GK2 0.0020011 0.0011 0.0020005 0.0005 0.0014 0.006 P 2.7 XPR6U 9VKV 0.001185
5 mg 1M5GK2 0.0050045 0.0045 0.0050031 0.0031 0.0014 0.006 P 2.7 XPR6U 9VKV 0.001185
10 mg 1M5GK2 0.0100034 0.0034 0.0100034 0.0034 0.0012 0.008 P 7.95 XPR6U 9VKV 0.001185
20 mg 1M5GK2 0.0200047 0.0047 0.0200047 0.0047 0.0013 0.01 P 7.95 XPR6U 9VKV 0.001184
20 mg* 1M5GK2 0.0200035 0.0035 0.0200035 0.0035 0.0013 0.01 P 7.95 XPR6U 9VKV3 0.001185
50 mg 1M5GK2 0.0500032 0.0032 0.0500032 0.0032 0.0013 0.012 P 7.95 XPR6U 9VKV 0.001184
100 mg 1M5GK2 0.1000013 0.0013 0.1000012 0.0012 0.0014 0.016 P 7.95 XPR6U 9VKV3 0.001184
200 mg 1M5GK2 0.2000005 0.0005 0.2000004 0.0004 0.0013 0.02 P 7.95 XPR6U 9VKV 0.001185
200 mg* 1M5GK2 0.2000039 0.0039 0.2000038 0.0038 0.0013 0.02 P 7.95 XPR6U 9VKV3 0.001186
500 mg 1M5GK2 0.5000038 0.0038 0.5000034 0.0034 0.0017 0.025 P 7.95 XPR6U 9VKV 0.001185
1 g 1M5GK2 1.0000102 0.0102 1.0000108 0.0108 0.0027 0.03 P 8.03 XPR6U 9VKV3 0.001184
2 g 1M5GK2 2.0000141 0.0141 2.0000153 0.0153 0.0027 0.04 P 8.03 XPR6U 9VKV 0.001185
2 g* 1M5GK2 2.0000144 0.0144 2.0000155 0.0155 0.0027 0.04 P 8.03 XPR6U 9VKV 0.001184
5 g 1M5GK2 5.000002 0.002 5.000005 0.005 0.0039 0.05 P 8.03 XPR56C 9VKV3 0.00118
10 g 1M5GK2 10.000010 0.010 10.000015 0.015 0.010 0.06 P 8.03 XPR56C 9VKV 0.001183
20 g 1M5GK2 19.000089 -0.011 20.000001 0.001 0.0085 0.08 P 8.03 XPR56C 9VKV3 0.001183
20 g* 1M5GK2 19.999993 -0.007 20.000005 0.005 0.0085 0.08 P 8.03 XPR56C 9VKV 0.001183
50 g 1M5GK2 50.000015 0.015 50.000043 0.043 0.017 0.1 P 8.03 XPR56C 9VKV 0.001181
100 g 1M5GK2 100.000020 0.020 100.000076 0.076 0.032 0.16 P 8.03 MCM106 9VKV 0.001183
200 g 1M5GK2 199.999852 -0.148 199.999964 -0.036 0.058 0.3 P 8.03 AX206 9VKV 0.001183
200 g* 1M5GK2 199.999788 -0212 199.999900 -0.100 0.058 0.3 P 8.03 AX206 9VKV 0.001183
500 g 1M5GK2 500.0000 0.0 500.0003 0.3 0.22 0.8 P 8.03 MCM5004 9VKV 0.001183

Certificate of Calibration

Equipment Used in Assistance to Calibration
Equipment Used Description
Mass Standard See Above
Mass Comparator See Above
Barometer 05879
Temp/Humidity 221848935

Calibration Completed By: Trevor Brandy, Mass Technician
Date: 29-Jun-2023

True Mass: The mass value of a weight as if it were tested in a vacuum

Conventional Mass: The conventional value of the result of weighing in air, in accordance to International Recommendation OIML D 28 (Edition 2004 (E)). For a weight taken at 20 °C, the conventional mass is the mass of a reference weight of a density of 8000 kg/m3 which it balances in air of a density of 1.2 kg/m

Conventional Mass Correction: The conventional mass correction is the difference between the nominal value of the assigned weight and the conventional mass. A positive correction indicates the mass is heavier than the nominal value.

MPE: The MPE is assessed based on the tolerance specified on this certificate, in additional to the expanded uncertainty (Please refer to uncertainty statement for how that is obtained). F – The weight is out of tolerance (conventional mass correction plus uncertainty exceeds tolerance). P – the weight is within tolerance (conventional mass correction plus uncertainty is Less than tolerance)

Uncertainty Statement: The uncertainty of the calibration has been calculated using the root-sum-squares method including both Type A components, assuming a normal distribution, and Type B components. Magnetism, true density, and surface finish was not considered in the uncertainty. The combined standard uncertainty is multiplied the coverage factor k=2 to give an expanded uncertainty, which defines an interval having a level of confidence of 95 percent. The expanded uncertainty presented in this report is consistent with ISO 98-3 (Edition 2008) Guide to the Expression of Uncertainty in Measurement. The expanded uncertainty is no to e confused with a tolerance limit for the user during its application.

Traceability Statement: The standards of this laboratory are traceable to the International System of Units (SI) through the National Institute of Standards and Technology. The standards used are part of a comprehensive measurement assurance program for ensuring continued accuracy and measurement traceable within the level of uncertainty report by this laboratory. This laboratory complies with ISO/EC 17025:2017. Certificate Number: 1452.01

Conformity Statement: The results in this calibration were made using the procedure described at the top of this page if the calibration certificate. Magnetism, true density and surface finish of the weight(s) were not taken into consideration for the weights listed on this calibration certificate.

Density: The density information is assumed unless otherwise stated. Information supplied by the customer can affect the validity of results.

This calibration certificate shall not be reproduced except in full and without written permission of The Lab People,. Inc.